Detailed product descriptionName:Closed Cup Flash Point TesterStandard:ASTM D93.Oil Cup Inner D
WhatsApp : +86 13816217984
Email : info@qinsun-lab.com
| Name: | Closed Cup Flash Point Tester | Standard: | ASTM D93. |
|---|---|---|---|
| Oil Cup Inner Diameter: | 50.7mm~50.8mm. | Oil Cup Depth: | 55.7mm~56.0mm |
| The Scribed Line Depth Of Capacity Of Testing Oil: | 33.9mm~34.3mm. | Capacity Of Testing Oil:: | About 70ml |
| Thermometers: | Scale 20°C~150°C, Division 1°C. | Keywords: | Oil Analysis Equipment |
| Highlight: | oil analysis equipmentoil analysis machine | ||
1.It adopts LCD screen to display. Full English man-machine dialog interface. It can preset the parameters of expected flash point, sample mark number, atmospheric pressure, test date, etc. It has menu to prompt and input function of guide type.
2.It adopts simulation tracking display the function curve of temperature rising and test time. It has prompt functions of English mis-operation, test date and test time.
3.It is equipped with standard RS-323.485 computer port. The lower computer can save 100 groups of historical data. The test data can be saved for a long time, transmitted and modified if the instrument being connected with computer.
4.It can correct the influence of atmospheric pressure and calculate the correction automatically.
5. Differential detection. Automatically correct the systematic deviation.
6. Automatically open the lid, ignite, detect and print test data. The test arm lifting up and down automatically.
7. It adopts electric ignition to fire light the gas flame. It will light only need to press the button.
8. Humanized design, beautiful and safe. Easy to operate.
1 .Power supply:AC(220±10%)V,50Hz
2. Temperature measurement Range: room temperature~300°C
Repeatability:≤3°C
Reproducibility:≤6°C
Accuracy:0.1°C
3. Heating rate: Procedure A:(5~6)°C/mi
Procedure B:(1~1.6)°C/min
Automatic control and manually adjustable.
4.Stirring rate
(1) Procedure A: (90~120)RPM
(2) Procedure B: (250±10)RPM
Automatic control and manually adjustable.
5.Igniting mode: Electric ignition. Gas flame diameter:3.2mm~4.8mm
6.Working condition: ambient temperature:(10~40)°C
relative humidity:≤80%
7. Maximum power consumption 500W
8. Dimension 520mm×360mm×310mm
| 1 | Power supply | AC(220±10%)V, 50Hz. |
| 2 | Flash point determination | Range: Ambient to 300℃. Meet requirement of GB/T 261-2008 Meet requirement of GB/T 261-2008 Resulotion: 0.1℃ |
| 3 | Heating rate | Procedure A: (5~6)℃/min Procedure B: (1~1.5)℃/min |
| 4 | Stirring rate | Procedure A: (90~120)RPM, Procedure B: (250±10)RPM Automatic control and manually adjustable. |
| 5 | Igniting mode | Igniting source: gas(or other civilian fuels)Electric ignition. Gas flame diameter: 3.2mm~4.8mm |
| 6 | Ambient temperature | ≤ 40℃ |
| 7 | Relative humidity | ≤ 80% |
| 8 | Maximum power consumption | 500W |
| 9 | Overall dimension | 410mm×360mm×310mm |
Qinsun Instruments Co., LTD is a professional laboratory testing instrument manufacturer in China,Have been focusing on laboratory instrument R&D more than 30 years and have rich industry experience,Based on international testing standards,We are also the instrument supplier for BV SGS laboratory,We provide one-stop solutions for lab instruments,Free Training and Turn-Key Service,Products exported all over the world,Offer 36 month warranty and are a trustworthy partner.