The light incubator is designed in accordance with the requirements of the “International Seed Inspection Regulations” of 1993 and manufactured in accordance with the relevant requirements of the national industry standard JB/T6823 “Technical Conditions for Bioartificial Climate Chambers”.
Biological culture testing device with four main functions: lighting, constant hot and cold temperature, automatic day and night switching program control and over-temperature protection.
Multi-segment programming control, computer control function, LCD display and other functions are optional.
It can be used for experiments such as germination of plant seeds, cultivation of seedlings, microbial culture, feeding of insects and small animals, etc.
Common Errors
The compressor is not cooling:
1. The compressor is always in equilibrium mode. When operating, the cooling effect is relatively good and the compressor is not workingnn be maintained effectively.
2. If it is a single standard system, it only works when the temperature drops to a certain value. When the door is opened and closed, there are large temperature fluctuations;
Cause Frequently starting the compressor can slightly shorten its lifespan.
Uneven temperature in the incubator:
1. This may be caused by discharging items too densely.
/ > 2. When the lamp is used for a long time, the bulb will heat up slightly, causing uneven temperatures
3. The circulation fan in the box does not work
Every refrigerated device has a defrost cycle in which the compressor does not cool.
Use the heating element to melt the ice on the condenser. The melted water can be behind the air outletaccumulate and cause freezing.
Solutions for common errors with light incubators
1. The cooling effect of the light incubator is poor or does not cool. What are the reasons:
First, you can open the glass door of the light incubator and put your hand on the bottom of the incubator to see if there is wind coming from the gap at the bottom; p>
If it says the fan is not broken, it is most likely a coolant problem or a clogged capillary tube. Otherwise, replace the fan.
2. The light incubator makes loud noises when it is not cooling. Why?
It may be that the circulating fan mounting screws are broken. The light incubator is loose. You can check the mounting screws. If they are loose, tighten them in time.
3. The lighting in the light incubator is not bright or dim. What should I do?
If theIf the lights are not lit, it is usually because the lamp legs are out of alignment, loose, or the electronic ballast or lamp tube is broken;
If the lights are dim, it may be because You only have a few lights on. It is recommended to turn on a few more lights.
The light incubator has an ultra-temperature sensor anomaly protection function to ensure the safety of instruments and samples; The optional full spectrum plant growth lamp has a positive effect on plant growth and improves disease resistance.
It has power-off memory and automatic power-off time compensation functions; a constant temperature control system, fast response and high accuracy of temperature control.
Microcomputer fully automatic control, touch switch, easy to useserve;
Programmable multi-stage control mode, the temperature, humidity and temperature can be adjusted Independent adjustment of day and night, lighting and time, etc.
Duct-type ventilation, gentle wind speed and uniform temperature in the studio; hollow reflective tempered glass with good thermal insulation performance and elegant appearance.
Fully enclosed, opaque lampshade, optional studio power supply, disinfection device, etc.