Mask synthetic blood penetration tester testing process

Mask synthetic blood penetration tester is used to measure the anti-splash penetration performance of protective masks such as medical masks, disposable masks, etc., and to evaluate the resistance of masks to the penetration of synthetic blood. Kong) Co., Ltd. Specializing in the production of synthetic mask blood penetration testers, we have been engaged in instrument testing for more than ten years. Welcome customers who have questions. MBMTextile Testing Instruments, Testing Equipment Manufacturer

Applicable Standards:MBMTextile Testing Instruments, Testing Equipment Manufacturer

IOS 19083-2010, YY/T 0469-2011, YY/T 0691-2008MBMTextile Testing Instruments, Testing Equipment Manufacturer

Test Process:

MBMTextile Testing Instruments, Testing Equipment Manufacturer

MBMTextile Testing Instruments, Testing Equipment Manufacturer

Test Steps: MBMTextile Testing Instruments, Testing Equipment Manufacturer

1. Pull the sample holder handle on the top left of the instrument and pull to the left. Lower the sample holder. Attach the sample (according to the standard, the sample must be tested within 60 seconds after it is removed from the pre-treatment chamber) on the sample holder. If the mask shows creases, unfold the creases and fix it on the sample holder to ensure that the target area is a single layer of material. The center of the sample was taken as the test target area. Close the specimen holder. MBMTextile Testing Instruments, Testing Equipment Manufacturer

2. Press the Start button. At this time, a certain amount of synthetic blood (surface tension (0.042 ± 0.002) N/m) is sprayed horizontally onto the mask under test behind the positioning target, and the spraying starts, and the time relay starts counting at the same time, and the instrument automatically stops spraying after the set injection time. MBMTextile Testing Instruments, Testing Equipment Manufacturer

3. After 10s the buzzer sounds,press the stop button to open the sample container. Under suitable lighting conditions, check

LabMBMTextile Testing Instruments, Testing Equipment Manufacturer

for synthetic blood on the observation surface of the sample. MBMTextile Testing Instruments, Testing Equipment Manufacturer

4. A test is over. If you need to continue the test, repeat steps 1-3 until the first part of the test is completed. MBMTextile Testing Instruments, Testing Equipment Manufacturer

5. After all tests are completed, clean up the synthetic blood residue in the test solution tank, replace it with distilled water, and clean the entire pipeline. Drain all water. Close the exhaust valve on the air compressor and turn off the power switch. Clean the residual liquid container, clean the work surface and remove the distilled water in the test solution box. MBMTextile Testing Instruments, Testing Equipment Manufacturer