Analysis of troubleshooting water bath shakers

Occasionally, while using the constant temperature shaker, errors such as: B. no display at startup, no rotation of the shaker plate, unstable vibration and loud device noise;pYBTextile Testing Instruments, Testing Equipment Manufacturer


How to deal with shaking in the water bath. Faults of the bed: pYBTextile Testing Instruments, Testing Equipment Manufacturer


1. The heating is normal, but the oscillation speed becomes slow. pYBTextile Testing Instruments, Testing Equipment Manufacturer


Treatment method: pYBTextile Testing Instruments, Testing Equipment Manufacturer


Open the side panel of the control cabinet, turn on the power and set the temperature controller to 0 degrees (i.e., no heating condition); pYBTextile Testing Instruments, Testing Equipment Manufacturer


Turn on the oscillation switch, adjust the speed to a higher level, locate the two adjustable resistors on the speed control board, and adjust the right resistor counterclockwise to solve the problem to solve the problem. pYBTextile Testing Instruments, Testing Equipment Manufacturer


2. The heating is normal, but the oscillation does not work. pYBTextile Testing Instruments, Testing Equipment Manufacturer


Treatment methode: pYBTextile Testing Instruments, Testing Equipment Manufacturer


Open the side panel of the control cabinet, turn on the power, set the temperature controller to 0 degrees (i.e. no heating state), turn on the oscillation switch;pYBTextile Testing Instruments, Testing Equipment Manufacturer


Adjust the speed to a higher level, use a multimeter to measure whether the transformer has a 220V voltage input at AC 250, if not, the switch is defective;

pYBTextile Testing Instruments, Testing Equipment Manufacturer


Measure conversely whether the transformer has a 12V voltage output (use AC 50V). Otherwise, the transformer is broken and replacing the transformer is not easy at this time ;pYBTextile Testing Instruments, Testing Equipment Manufacturer


This situation is mostly fixed in the circuit board. Some parts or motors have short-circuit failures (this phenomenon occurs with the water bath shaker, and the appearance of the transformer is burnt). If the above conditions are excluded, replace it. pYBTextile Testing Instruments, Testing Equipment Manufacturer


3. The display is normalal, the setting is normal and it can not be heated. pYBTextile Testing Instruments, Testing Equipment Manufacturer


Treatment method: pYBTextile Testing Instruments, Testing Equipment Manufacturer


(1) Turn on the power, turn on the power switch, and adjust the set temperature to be higher than the actual measured temperature and verify that there is an output indicator on the water bath shaker. If so, measure whether the heater tube has a voltage input. pYBTextile Testing Instruments, Testing Equipment Manufacturer


If the temperature controller has no voltage output indicator, it is recommended to replace the temperature controller. pYBTextile Testing Instruments, Testing Equipment Manufacturer


(2) The temperature controller displays LLL or 999 (depending on the manufacturer, the parameters are different);pYBTextile Testing Instruments, Testing Equipment Manufacturer


The failure of the water bath shaker is temperature If If the sensor is open or shorted, simply replace it. pYBTextile Testing Instruments, Testing Equipment Manufacturer