Next, we analyze the troubleshooting aspects of the high-temperature dust-free furnace:
1. Check the differential pressure gauge
Each high-temperature and dust-free furnace is equipped with a differential pressure gauge. In the periodic maintenance of the device, you need to check this table under the working condition of the device.
In order to properly judge whether the wind speed in the dust-free high-temperature furnace is normal.
If the differential pressure gauge value is found to be abnormal, check whether the heated objects are too dense, whether the filter is clogged, whether the air volume adjustment plate has been changed, etc.
2. Check the circulating fan.
The function of the circulating fan in the high temperature dust-free oven is to dissipate the heat, which can ensure that the air in the oven has some degree of mobility;
Hence the way it isWhen doing maintenance, we should also pay attention to whether the circulation fan is normal.
If there is an abnormality here, the hot air circulation in the high-temperature dust-free furnace is not good, so the hot and humid air inside cannot be discharged normally when the equipment breaks down.
3. Troubleshooting the furnace circuit
Because the high-temperature dust-free furnace is controlled by an automatic program, during routine maintenance and troubleshooting;
Be sure to inspect the circuit. If the high-temperature dust-free furnace can keep heating;
It means that there is no abnormality in the output signal of the intelligent computer temperature controller. On the contrary, it is necessary to check the various red line circuits.