The light incubator is designed in accordance with the requirements of the International Seed Inspection Regulations 1993 and manufactured in accordance with the relevant requirements of the national industrial standard JB/T6823 Technical Conditions for Biological Artificial Climate Chambers.
It can also be equipped with multi-segment programming controller, computer control function, liquid crystal display and other functions.
It can be used as a test for plant seed germination, seedling raising, microbial culture, feeding insects and small animals, etc.
Common mistakes
The compressor does not cool:
1 The compressor is always in balanceht mode The cooling effect in the working state is better, and the compressor can be maintained effectively.
2. If it is single system operation, it only works when the temperature drops to a certain level, and the temperature fluctuates greatly when the door is opened and closed.
Cause Frequent starting of the compressor tends to shorten its lifespan.
The temperature in the incubator is not uniform:
1. This may be caused by discharge objects that are too dense
2. If the light is used for a long time, the bulb will heat up easily, causing uneven temperatures
3. The circulation fan in the box is not working.
Recirculating air outlet Freezing:
Every refrigerator has a defrost cycle in which the compressor does not cool.
The heating element melts the ice on the condenser. The melted water canaccumulate behind the air outlet and lead to freezing.
Solutions to common mistakes of light incubator
1. The cooling effect of light incubator is not good or not, what is the reason:
If it does that the fan is not broken, there is probably a coolant problem or capillary clogging. Otherwise replace the fan.
2. The light incubator makes a lot of noise when not cooling. What is the reason?
Possibly the fixing screws of the light incubator circulation fan are loose. You can check the mounting screws. If they are loose, tighten them in time.
3. What should I do if the light of the light incubator is not on or is relatively dark?
The lights are not on, most of the lamp pins are misaligned, or the lamp pins are loose, or the electronic ballast or lamp tube is broken;
Because the lights are relative weak, you may only have a few lights on. It is recommended to turn on a few more lights.
The light incubator has an abnormal over-temperature sensor protection function to ensure the safety of the device and samples; the optional full-spectrum plant growth lamp promotes plant growth and improves disease resistance.
It has the function of power-off memory and automatic power-off time compensation; the constant temperature control system has fast response and high temperature control accuracy.
Microcomputer automatic control, touch switch, easy to operate;
Programmmable multi-segment control mode, temperature, humidity, lighting and time, etc.
Duct ventilation, studio gentle wind speed, temperature uniformity; hollow reflective tempered glass, good thermal insulation performance, beautiful and generous.
Fully enclosed opaque lampshade, optional studio power supply, disinfection device, etc.