Operation microcomputer control typeWhen using the universal electronic testing machine, there are sometimes software problems. The following are some software errors that may appear while using the electronic universal testing machine. You can fix them first following the above troubleshooting methods.
1, Online Command PromptError opening device, please check if the host is powered on or whether the communication line is properly connected!
Remedy: 1. Check the power indicator on the testing machine to see if it is on; 2. After startup, wait 10 seconds to connect; 3. Check whether the communication line between the testing machine and the computer is online; 4. Check whether the USB hardware driver is normal. If the driver is not normal, reconnect the USB plug and restart the test. If the driver still fails, please reinstall the driver software.
2, no curve display after the test runs
Error recovery method: 1. Check if curve coordinates exist. If noe curve coordinates are available, double-click on the result display bar. 2. Check the force, displacement, deflection, large deflection and other related information in the status display bar throughout the test process 3. Check whether the selection of the testing machine sensor and the coordinates of the curve plot are correct.
3, the device does not stop after sample is paused
Remedy: 1. Enter the fixed force (load) damping rate (amplitude) you specified in the test plan edit and set the test after you changed the value to a smaller value: 2. Increase the inlet force (load) and outlet load properly andmake sure the inlet load and outlet load settings are consistent.
4, stop the test if the test is not broken
Remedy: 1. Enter the Fixed Force (Load) Decay Rate (Amplitude) set in Edit Test Plan, change the value to a larger value and then test (if you do a peel or tear test, uncheck the previous one and do not need to adjust the damping); reduce the inlet force (load) and outlet load accordingly and ensure that the inlet load matches the outlet load setting 2. If not excluded, check in the test status display area to see which parameter the displayed switch-off condition is and modify it after confirming the parameter in the test plan.
5, Reuse Software Print Graphics orExport EXCELThe software displays an error message during the function
Remedy: When using these functions again, the software calls the default printer. Please check if there is a printer installed on the computer. If there is no printer, you can add a virtual printer.
6. During the experiment, it was noticed that the test results were quite different from the usual results
Remedy: 1. Please check whether the sensor selection in the Online setting is correct and the selection in the software fully matches the actual sensor model 2. Check whether the calibration or calibration function of the software has been used before the problem occurs Check whether the calibration value, the calibration value or any other information in the hardware parameters has been changed manually 3. Reboot the test machine Check host power supply and software if it is normal.
7 to run the software you mustYou will be prompted to enter the registration code or the trial period will expire(Timeout)
Troubleshooting method: The testing machine manufacturer's operating software is usually copyrighted, and this reason usually occurs after a reinstallation of the software or The manufacturer specifies a period of use, please contact contact the manufacturer's customer service in good time to request a registration code or password for lifetime use.
Reminder: If you encounter a problem that you cannot solve by yourself, it is normal to contact the manufacturer in time and avoid using other people's microcomputer control softwareMakers at will Connect to the universal electronic testing machine to avoid accidental damage to the instrument while the machine is running.