Salt Spray Corrosion Test Chamber Maintenance and Precautions

1. The maximum pressure of the pressure control valve on the back of the salt spray corrosion test chamber is 2kg, do not set the pressure too large, otherwise it may damage the pipe fittings. UPlTextile Testing Instruments, Testing Equipment Manufacturer

2. The maximum pressure of the pressure control valve in the front is set to Excessive pressure will damage the pipe fittings. UPlTextile Testing Instruments, Testing Equipment Manufacturer

3. The manual water switch must be turned off when refill water, otherwise the water in the pressure barrel will be spouted out from the manual water filling port. UPlTextile Testing Instruments, Testing Equipment Manufacturer

4. Pay attention to the indicator of the lack of water brine tank, if beforehanden If the brine water shortage indicator flashes, top up with brine immediately. UPlTextile Testing Instruments, Testing Equipment Manufacturer

5. Do not place the salt spray can outdoors, avoid direct sunlight. UPlTextile Testing Instruments, Testing Equipment Manufacturer

6. Handle it with care.UPlTextile Testing Instruments, Testing Equipment Manufacturer

7. When the spray nozzle is broken or the spray volume is insufficient ,UPlTextile Testing Instruments, Testing Equipment Manufacturer

① Please check whether the air pressure is normal; UPlTextile Testing Instruments, Testing Equipment Manufacturer

②Check whether the small holes at the front of the two glass tubes of the nozzle are on the same straight line; UPlTextile Testing Instruments, Testing Equipment Manufacturer

③Check whether the nozzle is clear. UPlTextile Testing Instruments, Testing Equipment Manufacturer

8. Please clean the salt spray chamber once or twice a month to keep the salt spray cleanchamber and reduce test errors.UPlTextile Testing Instruments, Testing Equipment Manufacturer

9. Place the fog collection funnel horizontally for accurate sampling.UPlTextile Testing Instruments, Testing Equipment Manufacturer

10 . If it is not used for a long time, please drain all the water in the salt spray chamber.

11. The touch screen and control box are precision electronic components. Prevent water, salt water or other liquids from getting inside.UPlTextile Testing Instruments, Testing Equipment Manufacturer